111
Specifications
33
1. Test with Analog Out (AN03)
Sample Rate
Process (100 mV)
TC (K type)
Rate (ms)
Rate (ms)
0
7(135)
7(135)
1
14(68)
14(68)
2
27(36)
27(36)
3
52(19)
41(24)
4
100(10)
41(24)
5
142(7)
41(24)
2. Test with Communication (Baud Rate at 19200 bps, Continuous Mode)
Sample Rate
Process (100mV)
TC (K type)
Rate (ms)
Rate (ms)
0
7(135)
7(135)
1
14(68)
14(68)
2
27(36)
27(36)
3
52(19)
33(30)
4
71(14)
33(30)
5
71(14)
33(30)
3. Test with Command Mode (Baud Rate at 19200 bps, Command Mode)
Sample Rate
Process (100mV) Rate (ms)
TC(K type) Rate (ms)
(Turn Around - Time ms)
(Turn Around - Time ms)
0
7(135 ms)
7(135 ms)
1
14(68 ms)
14(68 ms)
2
27(36 ms)
28(36 ms)
3
52(19 ms)
28(30 ms)
4
60(14 ms)
28(30 ms)
5
60(14 ms)
28(30 ms)
Table 33-1 Reading Rate
For the Command Mode, if the command-sending turn-around time is shorter than the
turn-around time in the above table (Refer to Process or TC), the response will have some
duplicate readings. For example, if the sample rate is 1 in process, the command-sending
turn-around time should be larger than 68 ms.
Содержание DP41-B Series
Страница 10: ...Notes viii...
Страница 53: ...43 Input Type Menu Flowchart 11 11 INPUT TYPE MENU FLOWCHART INPUt...