14
Reading Configuration Menu
52
Press ‘
RESET/ENTER
’
4) To enter Reading Scale & Offset configuration Submenu if display
shows
"
Rd.SC.OF
"
(First Item of
"
RdG.CNF
"
Menu).
Press ‘
RESET/ENTER
’
5) Display shows and flashes
"
dISAbL
"
disable (as default, Reading
Scale=1.0 and Reading Offset=0.0).
Press ‘
䊱
/MAX
’
6) Display flashes either
"
ENAbLE
"
or
"
dISAbL
"
. Select your choice and
press ‘
RESET/ENTER
’. If
"
dISAbL
"
selected, display advances to
"
UNt.tMP
"
Temperature Unit selection Submenu.
* If you select ENABLE, the following Steps 7 through 32 describe Reading Scale &
Offset Format Submenu. Otherwise continue with
"
UNt.tMP
"
in next section.
* For more detail, example, formula and Calculation Table, Refer to
Section 18
of
MULTI-POINT SCALE & OFFSET MENU CONFIGURATION
.
Enter READING SCALE & OFFSET FORMAT SUBMENU:
Press ‘
RESET/ENTER
’
7) To enter Reading Scale & Offset Format Configuration Submenu. Display
shows
"
StOREd
"
stored message momentarily then flashes
"dIRECt"
(default) Direct format or previous format selection, if your option of
"Rd.SC.OF"
or previously setup was
"
ENAbLE
"
.
Press ‘
䊱
/MAX
’
8) Scroll between two available formats:
"
dIRECt
"
or
"
2.Cd INt
"
(2-coordinate format) for the configuration format
of your choice. If your choice of selection is
"
2.Cd INt
"
, skip to Step 17).
dIRECt
READING SCALE & OFFSET DIRECT FORMAT MENU:
Direct scale and Offset: these two values are used in the straight line
equation, y = mx + b. (
"
Rd.SC.OF
"
):
Display Reading = m times input plus b
or Display Reading = [ m (input) + b ]
(Where m is the
"
RdG SC
"
and b is the
"
RdG OF
"
).
Press ‘
RESET/ENTER
’
9) Display still shows
"
dIRECt
"
and stops flashing.
Press ‘
RESET/ENTER
’
10) Display shows
"
RdG SC
"
Reading Scale.
Press ‘
RESET/ENTER
’
11) Display shows Direct Reading Scale Value (1.00000 as default or
previous setup value) with first digit flashing.
Press ‘
䊱
/MAX
’ & ‘
䊳
/MIN
’ 12) Use ‘
MAX
’ and ‘
MIN
’ buttons to enter new value of
"
RdG SC
"
Direct
Reading Scale.
Press ‘
RESET/ENTER
’
13) Display advances to
"
RdG OF
"
Direct Reading Offset. If a change was
made, the controller will also show
"
StOREd
"
stored message
momentarily.
Press ‘
RESET/ENTER
’
14) Display shows Direct Reading Offset Value (-000.00 as default or
previous setup value) with first digit flashing.
Press ‘
䊱
/MAX
’ & ‘
䊳
/MIN
’ 15) Use ‘
MAX
’ and ‘
MIN
’ buttons to enter new value of
"
RdG OF
"
Direct
Reading Offset.
Press ‘
RESET/ENTER
’
16) Display advances to
"
UNt.tMP
"
Temperature Unit Selection Menu. If a
change was made, the controller will also show
"
StOREd
"
stored message
momentarily.