18
Multi-Point Scale and Offset Menu Configuration
68
Determine the correct values for Inputs (
"
INPUt 0
"
and
"
INPUt 1
"
). Calculate
"
INPUt 0
"
and
"
INPUt 1
"
using the following equation:
"INPUt" = (Sensor Output) x (Conversion Number)
Conversion number is a coefficient of conversion between input values and real full display
range (100000 counts). See table below for proper conversion number.
INPUT RANGE CONVERSION NUMBER
0 – 100 mV
100000/(100 x 1)=1000cts/mV
0 - 1 V
100000/(1000 x 1)= 100cts/V
0 - 10 V
100000/(1000 x 10)= 10cts/V
0 – 20 mA
20000/ (20 x 1) = 1000cts/mA
Table 18-1. Conversion Number Table
Determine
"
INPUt 0
"
and
"
INPUt 1
"
Input range. For our transducer select 0 – 100 mV range.
"
INPUt 0
" = 0 (mV) x 1000(cts/mV) =
000000
"
INPUt 1
" = 30 (mV) x 1000(cts/mV) =
030000
1. Determine correct values for Display Reading (
"
REAd 0
"
and
"
REAd 1
"
)
2. In most cases,
"
REAd 0
"
and
"
REAd 1
"
are equal to the minimum and the maximum of the
transducer output range.
"REAd 0" = 000000
"REAd 1" = 000100
3. Scaling the controller.
Press ‘
RESET/ENTER
’
18) Display stops flashing
"
MANUAL
"
Manual Scaling method menu.
Press ‘
RESET/ENTER
’
19) Display shows
"
StOREd
"
momentarily and advances to
"
INPUt 0
"
heading.
Press ‘
RESET/ENTER
’
20) Display shows last stored
"
INPUt 0
"
value with 1st digit flashing.
Press ‘
䊱
/MAX
’ & ‘
䊳
/MIN
’ 21) To enter
"
INPUt 0
"
value (000000)
Press ‘
RESET/ENTER
’
22) Display advances to
"
REAd 0
"
heading.
Press ‘
RESET/ENTER
’
23) Display shows last stored
"
REAd 0
"
value with 1st digit flashing.
Press ‘
䊱
/MAX
’ & ‘
䊳
/MIN
’ 24) To enter
"
REAd 0
"
value (000000).
Press ‘
RESET/ENTER
’
25) Display shows
"
INPUt 1
"
heading.
Press ‘
RESET/ENTER
’
26) Display shows last stored
"
INPUt 1
"
value with 1st digit flashing.
Press ‘
䊱
/MAX
’ & ‘
䊳
/MIN
’ 27) To enter
"
INPUt 1
"
value (030000).
Press ‘
RESET/ENTER
’
28) Display advances to
"
REAd 1
"
heading.
Press ‘
RESET/ENTER
’
29) Display shows last stored
"
REAd 1
"
value with 1st digit flashing.
Press ‘
䊱
/MAX
’ & ‘
䊳
/MIN
’ 30) To enter
"
REAd 0
"
value (000100).
Press ‘
RESET/ENTER
’
31) Display shows
"
StOREd
"
momentarily and advances to
"
OUt.CNF
"
Output Configuration Menu.