95
Communication Menu Configuration
26
26.5
AddRES
Enter ADDRESS SELECTION SUBMENU
Press ‘
RESET/ENTER
’
60) Display shows
"
StOREd
"
stored message momentarily and then
advances to the
"
AddRES
”
Address selection submenu.
• This menu is applicable to the RS-485 Option only to designate different
address for each device or controller in the Multi-point Serial
Communication network that allows hosting device to recognize and
communicate properly as labeling or naming purposes.
Press ‘
RESET/ENTER
’
61) Display flashes
"
001
"
Address Value 1 as default.
Press ‘
䊱
/MAX
’ & ‘
䊳
/MIN
’ 62) To modify and select Address Value for your choice of setting. This
controller supports value within 000 – 199.
26.6.
SERCNt
Enter SERIAL COUNT SUBMENU
Press ‘
RESET/ENTER
’
63) Display shows
"
StOREd
"
stored message momentarily and then
advances to the
"
SERCNt
"
Serial Count submenu.
• Serial Count menu is used to program the number of readings the meter
just take (programmable from
"
00001
"
to
"
59999
"
) between
transmissions of data via the Serial Communication Option Board.
Caution:
For proper communication, do not set Serial Count Value at
"00000"
.
Press ‘
RESET/ENTER
’
64) Display flashes
"
00001
"
as default Serial Count value.
Press ‘
䊱
/MAX
’ & ‘
䊳
/MIN
’ 65) To modify and select the desired value of Serial Count (number of
readings) between data transmissions.
Press ‘
RESET/ENTER
’
66) Display shows
"
StOREd
"
stored message momentarily and then
advances to the
"
COLOR
"
LED Color Select submenu.