107
Troubleshooting
32
“I OVSC”
POSSIBLE CAUSE:
The input scale and/or offset values chosen are large enough to
drive the display into numerical overflow.
TO CORRECT:
Reduce the input and/or the input scaling/offset. Refer to
“
IN.SC.OF
”
in
Section 16.4
.
“R OVSC”
POSSIBLE CAUSE:
The reading scale and/or offset values chosen are large enough to
drive the display into numerical over flow.
TO CORRECT:
Reduce the reading scale/offset and/or move the active decimal
point to the right. Reducing input scale/offset is not re quired,
because the input overscale message has higher priority than this
message and would be displayed if there were an input overscale.
“CB OVF”
POSSIBLE CAUSE:
CNT BY (count by value) has been changed causing the display to
round up to a nu mer i cal overflow.
TO CORRECT:
Reduce the
“
CNT BY
”
count by value to 001. If you have an active
decimal point selected, move the decimal point one or more
positions to the right.
“SERIAL”
POSSIBLE CAUSE:
A configuring change has been attempted via the front panel
buttons while the serial communications port is actively
communicating with the meter.
TO CORRECT:
Either disconnect the serial cable, or press
‘MENU
’ for configuration
of the meter.