Test Programs
Maintenance Manual
3-22
Page
No. 15
Notch
alignment
adjustment
<Notch alignment adjustment>
This program can be used to check and adjust the notch positions f.
①
Insert a wafer into any slot of a cassette. and place the cassette on the cassette installation part.
②
Using the 4-way button, select the number of wafer type to adjust or register from the list of wafer
type numbers displayed on the liquid crystal panel, and then press the [Start] button.
(Registration is done for each wafer diameter. Registration for each wafer thickness is not
available.)
③
Mapping is performed and the [Wafer No.] button for a slot in which there is a wafer blinks.
Press the [Wafer No.] button to specify the slot into which the wafer to be adjusted has been
inserted. The [Wafer No.] button for the selected slot stops blinking and stays lit.
④
Press the [Start] button.
⑤
The selected wafer is loaded from the cassette.
⑥
Adjust the notch positions.
Data saved in the internal memory is displayed on the liquid crystal panel.
<CW/CCW 00>
⑦
Set the notch position setting knob in the direction you want to adjust the notch alignment.
⑧
Correct the notch alignment position. Save the correction data in the internal memory.
Use the 4-way button to correct the position.
4-way button [RIGHT]: Clockwise
4-way button [LEFT]: Counterclockwise
One pulse: About 0.1 degree
⑨
Register the correction value.
[M1]: Register <Save>
⑩
Press the [Start] button to repeat steps
to
.
•
If you do not press [M1] button but press the [Quit] button instead, the wafer is unloaded into the
cassette without saving the data.
•
Use the [Wafer alignment] control for adjusting the orientation flat positions.
•
Correction data must be registered separately for each direction of notch.
•
[M3]
:
Unload
•
[M4]
:
Load
No. 16
L-arm
rotation
origin
sensor
position
check
<
L-arm rotation origin sensor check >
This program can be used to check and adjust the L-arm rotation origin.
Press the [Start] button.
Select the item to adjust using the menu buttons on the liquid crystal panel.
[M1]
:
Check the L-arm rotation origin sensor
<L-arm rotation check>
[M2]
:
Check the L-arm wafer support unit <L-arm wafer support unit check>
Содержание AL120-12 Series
Страница 2: ...This page intentionally left blank...
Страница 13: ...Operation Maintenance Manual Introduction Conformity Standards i 11 Page...
Страница 14: ...Introduction Conformity Standards Operation Maintenance Manual i 12 Page...
Страница 16: ...Maintenance Manual Table of Contents C 2 Page This page intentionally left blank...
Страница 59: ...Maintenance Manual Test Programs 3 31 Page This page intentionally left blank...
Страница 67: ...This page intentionally left blank...