Maintenance Manual
Test Programs
3-11
Page
No. 7
L-arm
rotation
check
<L-arm rotation check>
This program can be used to check the L-arm rotation.
Press the [Start] button. If the loader has not been initialized, initialization starts.
Select the operation speed using the menu buttons on the liquid crystal panel.
[M1]: High speed <Fast speed> [M2]: Medium speed <Middle speed> [M3]: Low speed <Slow
speed>
•
If you do not specify anything here, the loader will operate at the high speed [M1]: High speed
<Fast speed> setting.
Press the [Start] button and the loader starts the operation.
①
Raises the L-arm
②
Rotates the L-arm to registered position
③
Rotates the L-arm when the rotation button is pressed
・
Press the button for registering the tilt position [Memory ], and the
L-arm rotates and stops at the registered position.
・
By pressing and holding down the button [Memory] for registering
the tilt position, you can register the position where the L-arm is
presently located.
By pressing the [M4]: <Return L-arm to registered tilt angle
position>, you can make the L-arm return to the initially registered
tilt position.
・
The speed can be changed during the pause.
・
When an error occurs, the loader displays the same error code
as in the regular operation, and then stops.
・
Press the [Quit] button, and the L-arm returns to its initial
position and the loader stops the operation.
To save the registered position, press the button [Memory] for
registering the tilt position.
Do not continue the operation for longer than 5 minutes.
Wait for longer than one minute before restarting the operation.
No. 8
F-arm
vertical
movement
check
<F-arm vertical movement check>
This program can be used to check the F-arm vertical movement.
Press the [Start] button.
Select the operation speed using the menu buttons on the liquid crystal panel.
[M1]: High speed <Fast speed> [M2]: Medium speed <Middle speed>
[M3]: Low speed <Slow speed>
・
If you do not specify anything here, the loader will operate at the high speed [M1]: High speed
<Fast speed> setting.
Press the [Start] button, and the loader starts the operation.
If the loader has not been initialized, initialization starts.
①
Raises the F-arm (Lower limit
→
Upper limit).
②
Lowers the F-
arm (Upper limit → Middle point).
③
Lowers the F-
arm (Middle point → Lower limit).
The loader then repeats steps
①
to
③
.
・
Press the [Start] button to pause the operation at the break point.
・
The speed can be changed during the pause.
・
When an error occurs, the loader displays the same error code as in the regular operation, and then
stops.
・
Press the [Quit] button, and the F-arm returns to its initial position and the loader stops the
operation.
CAUTION
Содержание AL120-12 Series
Страница 2: ...This page intentionally left blank...
Страница 13: ...Operation Maintenance Manual Introduction Conformity Standards i 11 Page...
Страница 14: ...Introduction Conformity Standards Operation Maintenance Manual i 12 Page...
Страница 16: ...Maintenance Manual Table of Contents C 2 Page This page intentionally left blank...
Страница 59: ...Maintenance Manual Test Programs 3 31 Page This page intentionally left blank...
Страница 67: ...This page intentionally left blank...