
Maintenance Manual
Test Programs
3-17
Page
No. 11
FOUP
Opener
mapping
parameter
setting
*Applicable
to
AL120-LMB1
2-LP3 only
<Mapping parameter check>
This program can be used to check and adjust the mapping.
Press the [Start] button, and select the item to adjust by using the menu buttons on the liquid crystal
panel.
[M1]
:
Automatically adjust mapping parameters <Mapping auto adjust>
[M2]
:
Execute mapping and display results <Mapping result display>
[M3]
:
Manually input mapping parameters <Input parameter>
[M1]: Automatically adjust mapping parameters <Mapping auto adjust>
The mapping parameters can be adjusted automatically.
Using the 4-way button, select the number of the wafer type to adjust or register from the list of wafer
type numbers displayed on the liquid crystal panel, and then press the [Start] button.
①
Insert the wafers to set into the lowest slot and the highest slot of the cassette. and place the
cassette on the cassette installation part.
②
Press the [M1] button, and press [START] button. The mapping is executed.
If the loader displays that the mapping data is <Result>: OK, press the [M2] button to register the data.
[M1]: Execute the automatic mapping adjustment <Adjust mapping>
[M2]: Register the results of automatic mapping adjustment <Save>
[M3]
:
Read default value from FOUP opener. <Load default parameter of FOUP opener>
[M4]:
Write default value to FOUP opener. <Set default parameter of FOUP opener>
* This test stores the standard values of the thickness, thickness tolerance and mapping start position
used for mapping the FOUP Opener. The mapping of the wafer to be set as the standard is executed
and the acquired value is set as the standard value.
* The position tolerance is fixed to 1.3mm. The thickness tolerance is 70% of the wafer thickness
acquired from FOUP Opener.
[M2]: Execute mapping and display results <Mapping result display>
This program can be used to check for normal detection when mapping is executed based on the
registered data.
Using the 4-way button, select the number of the wafer type to adjust or register from the list of wafer
type numbers displayed on the liquid crystal panel, and then press the [Start] button.
[M1]: Execute mapping <Mapping>
[M2]: Display results of mapping sensor <Display result >
[M1]: Execute mapping <Mapping>
①
Insert all wafers to be confirmed into the cassette. and place the cassette on the cassette
installation part.
②
Press the [M1] button. The mapping is executed.
[M2]: Display mapping results <Display result>
①
Press the [M2] buttons to display the results on the liquid crystal panel.
The results displayed include the results of judgments for each slot (Pass/Fail), center position of a
wafer <P>, and wafer thickness <Thick>.
[M1]: Next page <PageUp>
[M2]: Previous page <PageDn>
[M3]: Terminate display <Exit>
Содержание AL120-12 Series
Страница 2: ...This page intentionally left blank...
Страница 13: ...Operation Maintenance Manual Introduction Conformity Standards i 11 Page...
Страница 14: ...Introduction Conformity Standards Operation Maintenance Manual i 12 Page...
Страница 16: ...Maintenance Manual Table of Contents C 2 Page This page intentionally left blank...
Страница 59: ...Maintenance Manual Test Programs 3 31 Page This page intentionally left blank...
Страница 67: ...This page intentionally left blank...