137
automatically and display the collected dynamic waveform at the
waveform area. At this moment, the Sample button turns to the
Stop button.
NOTE: After creating new object, the software
creates a non-crack-across list by default. At this moment, the
New crack button at the main interface is unavailable. After
testing the non-crack-across data (under the condition that the
not across craks test points must be equal to or more than 3,
the regression calculation can be conducted), and the New
crack button is valiable.
1. Save the first test point
Press the Save button after adjusting the first pair of test point
waveform, save the first test point data and refresh the data list area
as well as display the saved test point data.
2. Test of the Subsequent Test Point
After the first test point having been saved, move the two
transducers down to the next test point, press the Save button upon
properly adjustment of waveform, repeat such process again and
again until the completing of test on all test points.
7.3.2 Stop sampling
During the sampling process, press the Stop button to stop