Novatest S.r.l.
Manual for Multichannel Ultrasonic Pile Integrity Tester
121
6.3 Start Test
6.3.1
Start sampling
Connecting two planar transducers with the transmitting channel
and receiving channel of instrument respectively by signal wire, and
couple transducers respectively on the first test point of the first test
area on the two test surfaces of the object to be tested, and press the
sampling button on the main interface, and then continuous sampling
will start automatically, and dynamic waveform will collected will be
displayed in the waveform area. At the moment the Sample button
will be changed into Stop button, Exit button became Search button
that will be used for searching head wave.
If the data is to be saved, the new object name shall be entered
prior to sampling; if no object name is entered, the Save button is
invalid.
1. Save the First Test Point
After the first test point having been adjusted, press the Save
button, and the current test point data will be saved and the data list
area will display one line of data saved a moment ago
2. Test of subsequent test points
After the first test point having been saved, move the two
transducers down to the next test point, press the Save button upon
properly adjustment of waveform, repeat such process again and
again until the completing of test on all test points.