11.2.4
Registering Calibration Settings
Calibration values to be reference lengths when displaying scales and making measurement can be registered on the [TOOL
SET: CALIB] screen.
Registering calibration settings
The DS-L3 has two types of calibration registration: [OPTICAL] and [MANUAL].
•
[OPTICAL]
This type allows you to input the magnification of optical equipment such as microscope. The DS-L3
calculates a calibration value from the entered value.
•
[MANUAL]
This type allows you to register an arbitrary calibration value by displaying an object of known length
and specifying the length on the screen.
Required tools
Prepare the following tools for registration using the [MANUAL] setting.
•
Object of known length
How to register
Follow the procedures described below to register user-configured calibration.
Selecting a Mode
[OPTICAL] mode
1
Display the [TOOL SET: CALIB] screen.
Press the [TOOL SET] button and press the [CALIB] button
on the submenu.
2
Select [MANUAL] or [OPTICAL] from the [MODE] area.
3
Perform either of the following operations according to
the selected mode.
When [OPTICAL] is selected with DS-L3 and the
microscope not connected with a USB cable
1. Enter the magnification of the objective of the
optical equipment in the input field of the
[OBJECTIVE] area.
2. If a relay lens is in use, enter the magnification of
the relay lens in the input field of the [RELAY] area.
When no relay lens is used, enter “1.00”.
When [OPTICAL] is selected with DS-L3 and the
motorized microscope connected with a USB cable
1. Enter the magnification of the relay lens in the input
field of the [RELAY] area.
Magnification is shown in the input field of the
[OBJECTIVE] area based on the microscope
information configured in advance. This value cannot be
edited here.
This magnification information is also registered to the
setting of calibration for a nosepiece with the matching
number.
The [INFO] area is hidden as the name of the
microscope connected to the system is automatically
registered as a comment.
Notes:
•
If the AZ100M microscope is connected to
DS-L3, upon detection of AZ-ICI coaxial
illuminator attached to the microscope the
system automatically performs adjustment
by multiplying the objective magnification by
1.25 to determine a measurement value.
For the [RELAY] area, you only need to
enter the magnification (0.6x) of the vertical
tube on AZ100M, but not the magnification
for the coaxial illuminator.
•
When the MA200 microscope is connected
to DS-L3, be sure that the magnification
module turret is set at a click-stop position.
Otherwise, the registration button cannot be
selected.
1
2
Select a mode.
3-1, 2
Enter the optical
equipment magnification.