![NI PXI-4204 Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/ni/pxi-4204/pxi-4204_user-manual_1685515069.webp)
©
National Instruments Corporation
D-1
D
Common Questions
This appendix lists common questions related to the use of the
NI PXI-4204.
Which version of NI-DAQ works with the NI PXI-4204 and how do
I get the most current version of NI-DAQ?
You must have NI-DAQ 7.0 or later and use NI-DAQmx. Visit
ni.com
and
follow the link,
Download Software»Drivers and Updates»Search
Drivers and Updates
, and enter the keyword
NI-DAQ
to find the latest
version of NI-DAQ for your operating system.
Does the NI PXI-4204 have hardware analog triggering?
No, the NI PXI-4204 does not support hardware analog triggering. The
NI PXI-4204 does support digital triggering.
Is the NI PXI-4204 an isolated device?
No, the NI PXI-4204 is not an isolated device and should not be used in
applications with working voltages larger than ±100 V. The NI PXI-4204
has only resistive isolation, which does not fit all definitions for isolation.
Whereas the NI PXI-4204 has a very high input range, the maximum
working voltage specification should be carefully observed. The balanced
differential inputs of the NI PXI-4204 provide effective rejection of
common-mode voltages resulting from ground loops. However, high CMV
can damage the NI PXI-4204. Refer to Appendix A,
maximum working voltage specifications. NI does offer signal
conditioning devices that provide high voltage isolation. For more
information about isolated NI products, visit
ni.com/sigcon
.
What is the function of the chassis ground screw?
The chassis ground screw is intended to provide a ground for shielded field
wiring.