Chapter 2
Connecting Signals
©
National Instruments Corporation
2-7
Shielded Floating Signal Source Connection
(Recommended)
The signal source shown in Figure 2-5 is identical to the one shown in
Figure 2-4. The only difference is the addition of shielding around the field
wiring.
You can improve the noise susceptibility of the floating signal source
connection, shown in Figure 2-4, by using a shielded cable to connect
the signal source to the instrument, as shown in Figure 2-5. This is the
recommended shielding configuration for floating signal sources. The
shielding is grounded at only one point to the NI PXI-4204 chassis ground
reference. Connect the signal (
V
SIG
+) to the NI PXI-4204 channel (AI
X
+).
Connect the signal reference (
V
SIG
–) to the channel reference (AI
X
–).
Figure 2-5.
Shielded Floating Signal Source Connection (Recommended)
Shielding
+
–
+
–
+
–
Twisted-Pair
Wiring
AI0+
AI0–
AI7+
AI7–
CH0
CH7
V
SIG
Chassis Ground
Reference
Signal Source