Appendix A
Specifications
©
National Instruments Corporation
A-3
NI PXI-4204 Accuracy Information with SS/H Disabled
NI PXI-4204 Accuracy Information with SS/H Enabled
SS/H Characteristics
Droop rate .............................................. <0.4 mV/ms
Hold mode settle time ............................ 3
μ
s min
Acquisition time ..................................... 7
μ
s min
Hold step ................................................ 20 mV
1
Nominal Range (V)
Absolute Accuracy
Absolute
Accuracy
at Full
Scale
(mV)
% of Reading
Offset
(mV)
Noise + Quantization
(mV)
Temp
Drift
(%/
°
C)
Positive
FS
Negative
FS
24 Hours
1 Year
Single
Pt.
Avg.
100
–100
0.063
0.070
±
16.8
±
9.34
±
0.92
0.0027
88
50
–50
0.031
0.038
±
10.4
±
4.68
±
0.57
0.0025
30
5
–5
0.061
0.068
±
4.7
±
0.69
±
0.40
0.0027
8.5
0.5
–0.5
0.061
0.068
±
4.1
±
0.49
±
0.40
0.0027
4.8
Accuracies are valid for measurements following an internal self-calibration. Averaged numbers assume dithering and averaging of
100 single-channel readings.
Measurement accuracies are listed for operational temperatures within ±1 °C of internal calibration temperature and ±10 °C of external or
factory calibration temperature.
Nominal Range (V)
Absolute Accuracy
Absolute
Accuracy
at Full
Scale
(mV)
% of Reading
Offset
(mV)
Noise + Quantization
(mV)
Temp
Drift
(%/
°
C)
Positive
FS
Negative
FS
24 Hours
1 Year
Single
Pt.
Avg.
100
–100
0.085
0.090
±
19.0
±
9.35
±
1.02
0.0027
110
50
–50
0.035
0.042
±
12.6
±
4.70
±
0.73
0.0025
34
5
–5
0.063
0.070
±
6.8
±
0.82
±
0.60
0.0027
11
0.5
–0.5
0.063
0.070
±
6.3
±
0.66
±
0.60
0.0027
7.2
Accuracies are valid for measurements following an internal self-calibration. Averaged numbers assume dithering and averaging of
100 single-channel readings.
Measurement accuracies are listed for operational temperatures within ±1 °C of internal calibration temperature and ±10 °C of external or
factory calibration temperature.
1
The 20 mV hold step error is calibrated out during an external calibration.