Chapter 2
Connecting Signals
©
National Instruments Corporation
2-11
Figure 2-8.
Alternative Shielded Ground Referenced Signal Connection
High CMV Connection
Figure 2-9 shows a signal source with a high CMV present. With this
configuration, the signal rides on an offset voltage called the CMV. The
CMV is not of interest to the measurement and can appear as an error in
the voltage measurement. This type of connection can be treated as a
ground-referenced signal source with a large difference between the signal
source reference and the NI PXI-4204 chassis ground reference.
To connect a signal source with high CMV to the NI PXI-4204, connect the
signal (
V
SIG
+) to the NI PXI-4204 channel (AI
X
+). Connect the signal
reference (
V
SIG
–) to the channel reference (AI
X
–).
The balanced front end of the NI PXI-4204 rejects CMV, making it possible
to measure signal sources in the presence of a large CMV. The only
constraint for the NI PXI-4204 is that the combined signal and CMV must
not exceed ±100 V. The maximum allowed combination of signal voltage
and CMV is specified as the maximum working voltage.
Shielding
+
–
+
–
+
–
Twisted-Pair
Wiring
AI0+
AI0–
AI7+
AI7–
CH0
CH7
V
SIG
V
SIG
Ground
Reference
Chassis Ground
Reference
Signal Source