![NI PXI-4204 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/ni/pxi-4204/pxi-4204_user-manual_1685515058.webp)
Appendix B
Timing Signal Information
B-2
ni.com
DAQ Timing Connections
The timing signals are AI START TRIG, AI REF TRIG, AI SAMP CLK,
AI CONV CLK, AI PAUSE TRIG, AI SAMPLE CLK TIMEBASE, and
AI HOLD COMPLETE.
Posttriggered DAQ allows you to view data that is acquired after a trigger
event is received. Figure B-1 shows a typical posttriggered sequence.
Figure B-1.
Typical Posttriggered Sequence
Pretriggered DAQ allows you to view data that is acquired before the
trigger of interest in addition to data acquired after the trigger. Figure B-2
shows a typical pretriggered sequence.
Figure B-2.
Typical Pretriggered Sequence
AI START TRIG Signal
The AI START TRIG signal can be input through PFI0, PXI_Trig<0..5>,
or PXI Star. The AI START TRIG signal can be output through PFI0 or
PXI_Trig<0..5>.
1
3
0
4
2
AI START TRIG
AI SAMP CLK
AI CONV CLK
Scan Counter
n/a
0
1
2
3
1
0
2
2
2
AI START TRIG
AI REF TRIG
AI SAMP CLK
AI CONV CLK
Scan Counter