Chapter 2
Connecting Signals
2-6
ni.com
Floating Signal Source Connection
Figure 2-4 illustrates the floating signal source connection. In this
configuration, the signal source being measured is a floating signal source,
such as a battery.
To connect a floating signal source connection to the NI PXI-4204, the
signal (
V
SIG
+) is connected to the NI PXI-4204 channel (AI
X
+). The signal
reference (
V
SIG
–) is connected to the channel reference (AI
X
–). The lack of
shielding on the field wiring can cause this configuration to be susceptible
to electrically coupled noise.
This configuration has only one ground connection—at the instrument—so
there is no potential difference between grounds to introduce error in your
measurements. The NI PXI-4204 input attenuators provide a ground
reference for floating signal sources; therefore, external grounding bias
resistors are not required.
Figure 2-4.
Floating Signal Source Connection
+
–
+
–
+
–
Twisted-Pair
Wiring
AI0+
AI0–
AI7+
AI7–
CH0
CH7
V
SIG
Chassis Ground
Reference
Signal Source