Chapter 2
Connecting Signals
©
National Instruments Corporation
2-9
Figure 2-6.
Ground-Referenced Signal Connection
Shielded Ground-Referenced Signal Connection
(Recommended)
Figure 2-7 illustrates a shielded ground-referenced signal connection. The
connection to this signal source is identical to the ground-referenced signal
connection with the addition of shielding around the field wiring.
The ground-referenced signal source in Figure 2-7 is identical to the signal
source in Figure 2-6.
The shielding is grounded at only one point to the signal source ground
(
V
SIG
Ground Reference). Connect the signal (
V
SIG
+) to the NI PXI-4204
channel (AI
X
+). Connect the signal reference (
V
SIG
–) to the channel
reference (AI
X
–).
This shielding scheme is effective at reducing capacitive or electrically
coupled noise. The same concerns regarding the difference in ground
potentials, discussed in the
Ground-Referenced Signal Connection
section,
also apply to this configuration.
+
–
+
–
+
–
Twisted-Pair
Wiring
AI0+
AI0–
AI7+
AI7–
CH0
CH7
V
SIG
V
SIG
Ground
Reference
Chassis Ground
Reference
Signal Source