Appendix B
Timing Signal Information
©
National Instruments Corporation
B-5
Figure B-6.
AI REF TRIG Output Signal Timing
AI SAMP CLK Signal
The AI SAMP CLK signal can be externally input from PFI0,
PXI_Trig<0..5>, or PXI Star. It can be output on any PXI trigger bus line.
Refer to Figures B-1 and B-2 for the relationship of AI SAMP CLK to the
sequence.
As an input, AI SAMP CLK is configured in edge-detection mode. You
can configure the polarity selection for either rising or falling edge.
The selected edge of AI SAMP CLK initiates a scan. The SI2 counter starts
if you select an internally triggered AI CONV CLK.
As an output, AI SAMP CLK reflects the actual start pulse that initiates
a scan, even if the starts are externally triggered by another PFI or
PXI_Trig<0..5>. You have two output options. The first option is an active
high pulse with a pulse width of 50 to 100 ns, which indicates the start of
the scan. The second option is an active high pulse that terminates at the
start of the last conversion in the scan, which indicates a scan in progress.
AI SAMP CLK is deasserted, t
off
, after the last conversion in the scan
is initiated. This output is set to high-impedance at startup.
t
w
= 50 to 100 ns
t
w