Chapter 2
Connecting Signals
2-10
ni.com
Figure 2-7.
Recommended Shielded Ground-Referenced Signal Connection
Alternative Shielded Ground-Referenced Connection
Figure 2-8 illustrates an alternative shielded ground-referenced signal
connection. The ground-referenced signal source in Figure 2-8 is identical
to the signal source in Figure 2-7. The only difference is where you wire the
shielded ground connection.
The shielded ground connection in Figure 2-7 is the most effective
configuration for reducing electrically induced noise in a
ground-referenced measurement. However, if connecting the shielding
to the signal source ground is cumbersome or inconvenient, use the
connection shown in Figure 2-8 to reduce electrically coupled noise. The
shielding is grounded at only one point to the NI PXI-4204 chassis ground
reference. Connect the signal (
V
SIG
+) to the NI PXI-4204 channel (AI
X
+).
Connect the signal reference (
V
SIG
–) to the channel reference (AI
X
–).
The same concerns regarding the difference in ground potentials, discussed
in the
Ground-Referenced Signal Connection
section, also apply to this
configuration.
Shielding
+
–
+
–
+
–
Twisted-Pair
Wiring
AI0+
AI0–
AI7+
AI7–
CH0
CH7
V
SIG
V
SIG
Ground
Reference
Chassis Ground
Reference
Signal Source