Appendix A
Specifications
©
National Instruments Corporation
A-7
Maximum Working Voltage
( common-mode) each input should remain within ±100 V of
ground.
Maximum working voltage refers to the signal voltage plus the CMV.
Channel-to-earth (inputs) ....................... 100 V, 70 V
rms
,
Measurement Category I
Channel-to-channel (inputs)................... 100 V, 70 V
rms
,
Measurement Category I
Cautions
This device is rated for Measurement Category I and intended to carry signal
voltages no greater than 100 V. This device can withstand up to 500 V impulse voltage
without creating a safety hazard. Do not use this device for connection to signals or for
measurements within Categories II, III, or IV. Do not connect to MAINS supply circuits,
such as wall outlets, of 115 or 230 VAC.
When hazardous voltages (>42.4 V
pk
/60 V) are present on any terminal, safety low-voltage
(
≤
42.4 V
pk
/60 V) cannot be connected to any other terminal.
Overvoltage Protection
AI0 to AI07 ............................................ ±110 V, powered on or off
SMB connector ...................................... ±15 V, powered on or off
Environmental
Operating temperature............................ 0 to 55 °C
Storage temperature ............................... –20 to 70 °C
Humidity ................................................ 10 to 90% RH, noncondensing
Maximum altitude .................................. 2,000 m
Pollution Degree (indoor use only) ........ 2