Chapter 3
Using the NI PXI-4204
©
National Instruments Corporation
3-3
Signal Conditioning Functional Overview
The NI PXI-4204 is part of the NI PXI-4200 series of DAQ devices
designed to provide application-specific signal conditioning, DAQ, and
integrated field wiring connectivity on the same product. The NI PXI-4204
signal conditioning circuitry is designed to provide attenuation,
amplification, filtering, and SS/H capability as described in Table 3-1.
Table 3-1.
Signal Conditioning Functional Blocks
Signal Conditioning Component
Description
Divide-by-Ten Attenuator
Each NI PXI-4204 channel front end has a balanced
divide-by-ten attenuator. The attenuation is provided by
a resistor-divider network that provides 1 M
Ω
input
impedance. The balanced attenuator followed by an
instrumentation amplifier provides rejection of
common-mode signals and enables the NI PXI-4204 to
measure voltages up to ±100 V.
Input Multiplexer
Each channel of the NI PXI-4204 includes an analog
multiplexer. You can use this multiplexer to
programmatically route internal test signals to the
NI PXI-4204 instrumentation amplifiers for self-calibration
and self-test.
Instrumentation Amplifier
The instrumentation amplifier following the attenuation stage
of the NI PXI-4204 provides a very high-load impedance to
the previous stage, rejection of common-mode signals, and
converts differential signals to single-ended signals that are
then fed to the filter stage.