4.3 Sorting the Part (Comparator)
4-8
ZM2371/ZM2372
Setting the Comparator
The comparator function is set with the comparator setting menu.
_
SHIFT
_
+ [ COMPRTR ]
Press the
_
SHIFT
_
+ [COMPRTR]
keys to display the comparator setting menu.
When limit comparison,
L/U
Comparator function
ON/OFF
Comparison format of primary parameter
C o m pa r a t o r : O F F
A B S
Current setting
0 ) O F F 1 ) O N 2 ) D E V
> N E X T
Options (first page)
3 ) C L E A R 4 ) L I M I T
< P R E V
Options (second page)
OFF
Disables the comparator function (Bin sorting). (Initial value)
ON
Enables the comparator function (Bin sorting).
DEV
Displays the primary parameter deviation comparison menu for
comparator.
CLEAR
Displays the comparator initializing menu.
LIMIT
Displays the comparator upper/lower limit setting menu.
L/U
Appears when primary or secondary parameter comparison function is
enables. The limit comparison function cannot be made enabled through
the panel. Turning on/off of comparator function disables the limit
comparison function of primary/secondary parameters.
If a setting is made or the EXIT operation is performed, the measurement screen comes
back.
COMPRTR
lamp
When the comparator function is enabled, the COMPRTR lamp on front panel lights up. It
also lights up when the limit comparison function is enabled.
COMPRTR lamp
_
SHIFT
_
_
4
_
| [COMPRTR]
_
5
_
| [Pri/Sec]
_
1
_
| [NO LIMIT]
_
EXP
_
| [SYSTEM]
Содержание ZM2371
Страница 1: ...NF Corporation LCR METER ZM2371 ZM2372 Instruction Manual...
Страница 2: ......
Страница 3: ...LCR METER ZM2371 ZM2372 Instruction Manual DA00026021 007...
Страница 250: ......
Страница 252: ......
Страница 253: ......
Страница 254: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...
Страница 255: ...BLANK...