3.5 Basic Operations
3-35
ZM2371/ZM2372
3.5.6.3 Measurement Range
Set the measurement range according to the impedance of DUT.
Current measurement range setting is displayed at the lower left of measurement screen.
・ ・ ・ ・ ・ ・ ・
A 1 0 0
F A S T 2 5 6
I n t O p S h L d 1 m
・ ・ ・ ・ ・ ・
RANGE
A
Auto: Automatic selection of measurement range
H
Hold: Measurement range is held for manual selection
100
Measurement range (100m
to 1M
)
Table 3
–2 Measurement range list
Measurement
range
Recommended
range
Measurable range
Remarks
1M
1M
to
11M
900k
to
frequency
≦
20kHz (*1)
100k
100k
to 1.1M
90k
to
-
10k
10k
to 110k
9k
to
-
1k
1k
to
11k
900
to
-
100
9
to
1.1k
Not limited
-
10
0.9
to
10
to 11
Output impedance 100
/5
1
90m
to
1.0
to 1.1
Output impedance 25
/5
100m
9m
to 100m
to 110m
Output impedance 25
/5
Recommended range: Recommended operating range for high accuracy measurement.
If out of the recommended range extremely, the measured value or data output may be
overflow.
Measureable range: Approximate range in which the measurement is possible.
The measurement may be possible even outside the range given above depending on the
measurement conditions, or the measurement may not be possible even within the
range given above due to incoming noise.
*1 When frequency > 20kHz, the recommended range, measureable range, and
meaurement accuracy are same as 100k
range (Zr = 100k
) even if 1M
range is
specified.
Table 3
–3 Max. current, max. voltage and output impedance in each measurement range
Measurement
range
Max. drive / detection
current
Max. voltage
Output impedance Rd
1M
5µArms
5Vrms
100
100k
50µArms
5Vrms
100
10k
500µArms
5Vrms
100
1k
5mArms
5Vrms
100
100
50mArms
5Vrms
100
10
200mArms
5Vrms
(1Vrms)
100
(5
)
1
200mArms
5Vrms
(1Vrms)
25
(5
)
100m
200mArms
5Vrms
(1Vrms)
25
(5
)
The values in ( ) are when the minimum output impedance is 5
, and the conditions for 5
are fulfilled. Allowable current peak value (instantaneous value) including DC bias current
is about 1.4 times the current value given in table (root-mean-square value).
Содержание ZM2371
Страница 1: ...NF Corporation LCR METER ZM2371 ZM2372 Instruction Manual...
Страница 2: ......
Страница 3: ...LCR METER ZM2371 ZM2372 Instruction Manual DA00026021 007...
Страница 250: ......
Страница 252: ......
Страница 253: ......
Страница 254: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...
Страница 255: ...BLANK...