3-14
ZM2371/ZM2372
Measurement signals
Main specifications of measurement signals are as follows:
Signal level
Voltage:
10mVrms to 5Vrms (ALC possible)
Constant current: 1
Arms to 200mArms (range depends on
measurement range, etc.)
Max. about 7.1Vpk voltage including DC bias is output.
Output impedance
About 5
/ 25
/ 100
(depending on measurement range, etc.)
Max. drive current
About 200mArms
3.4.2 Connection to DUT
Connect
ZM2371 / ZM2372
and DUT with 4-terminal (4-terminal-pair) as shown below to
avoid the influence of contact resistance.
Figure 3
–4 Connection to DUT
In the measurement of high impedance, shielding around the DUT can restrict variation of
measured value. In a simplified method, place an insulating plate on the top surface of
ZM2371 / ZM2372
, and measure the DUT on it.
The
ZM2371 / ZM2372
cannot measure the grounded DUT. Both ends of DUT must be both
insulated from the grounding.
When commercially available test fixture or test lead is used, refer to the instruction
manual of these products. The product in which the shields of respective terminals are not
connected cannot be used together with the
ZM2371 / ZM2372
.
DUT
H
CUR
L
CUR
H
POT
L
POT
Shield plate
Insulating plate
LCR meter
Use four coaxial cables having BNC
connector.
Connect 4 shields together on the
DUT side.
When the measured value is unstable due to a noise, lay a metallic plate connected
to the outer conductor (shields) of measurement cables or the case under the DUT
for shielding.
Содержание ZM2371
Страница 1: ...NF Corporation LCR METER ZM2371 ZM2372 Instruction Manual...
Страница 2: ......
Страница 3: ...LCR METER ZM2371 ZM2372 Instruction Manual DA00026021 007...
Страница 250: ......
Страница 252: ......
Страница 253: ......
Страница 254: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...
Страница 255: ...BLANK...