5.6 Command Explanation
5-34
ZM2371/ZM2372
The primary parameter limit comparison fail flag which is a response
of :CALCulate1:LIMit:FAIL? query is updated when a new measured value is obtained. At
this time, it is assumed as pass if the limit comparison function of primary parameters is set
to OFF (disabled). This is also true for the secondary parameters. The limit comparison fail
flag is initialized to “0” (pass) when the power is turned on. It can also be initialized by the
*RST command, initialize operation with the initialize menu or system setting menu,
or :CALCulate{1|2}:LIMit:CLEar command.
:CALCulate:FORMat:AUTO[:STATe] {ON|OFF|1|0}
:CALCulate:FORMat:AUTO[:STATe]?
Description:
Sets/queries whether primary and secondary measurement parameters
are selected automatically.
Parameters:
{ON|1} Enables the automatic selection of measurement parameter.
{OFF|0} Disables the automatic selection of measurement parameter.
Setting example:
:CALC:FORM:AUTO ON
Sets to enables the automatic selection of measurement parameter.
Response:
{1|0}
Query example:
:CALC:FORM:AUTO?
Response example: 1
The automatic selection of parameter is enabled.
Remarks:
Automatic selection of measurement parameters is automatically
disabled in the following cases:
•
When a primary parameter is set
•
When a secondary parameter is set
•
When automatic selection of equivalent circuit is set to OFF
(disabled)
•
When a setting is made with :SENSe:FUNCtion command
:CALCulate1:CKIT:AUTO[:STATe] {ON|OFF|1|0}
:CALCulate1:CKIT:AUTO[:STATe]?
Description:
Sets/queries whether equivalent circuit is selected automatically when
the primary parameter is set to either R, C, or L.
Parameters:
{ON|1} Enables the automatic selection of equivalent circuit.
{OFF|0} Disables the automatic selection of equivalent circuit.
Setting example:
:CALC1:CKIT ON
Sets to enables the automatic selection of equivalent circuit.
Response:
{1|0}
Query example:
:CALC1:CKIT?
Response example: 1
The automatic selection of equivalent circuit is enabled.
Содержание ZM2371
Страница 1: ...NF Corporation LCR METER ZM2371 ZM2372 Instruction Manual...
Страница 2: ......
Страница 3: ...LCR METER ZM2371 ZM2372 Instruction Manual DA00026021 007...
Страница 250: ......
Страница 252: ......
Страница 253: ......
Страница 254: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...
Страница 255: ...BLANK...