5.6 Command Explanation
5-54
ZM2371/ZM2372
:INITiate:CONTinuous {ON|OFF|1|0}
:INITiate:CONTinuous?
Description:
Sets/queries whether the trigger system continuous initiation is enabled
or disabled.
If enabled, when the measurement finished and the trigger system
becomes idle state, the device goes in the trigger waiting state
automatically. If disabled, the device remains in idle state.
Setting example:
:INIT:CONT ON
Enables the trigger system continuous initiation.
Response:
{1|0}
Query example:
:INIT:CONT?
Response example: 1
The trigger system continuous initiation is enabled.
Remarks:
The continuous initiation is enabled when the power is turned on.
It is also enabled when the initialization is executed by panel operation.
The continuous initiation is disabled by the *RST command.
:INITiate[:IMMediate]
Description:
The device goes in the trigger waiting state when the trigger system is
in idle state.
Setting example:
:INIT
Sets the trigger waiting state.
:READ?
Description:
The measurement is stopped and the device goes in the trigger waiting
state whichever state the device is placed in at that time. After that,
when the trigger is applied and one-time measurement finished, the
measured data is read.
After :READ? query is transmitted, receive the measured data. Upon
completion of the measurement, newly measured data can be received.
Response:
<measurement status>, <primary parameter measured value>,
<secondary parameter measured value>
[, { <comparator’s bin sorting result>|<limit comparison result> } ]
Query example:
:READ?
Response example: +0,+3.1415E-06,+1.20000,+2
The response message is same as :FETCh? query. For details, see the
description of :FETCh? query.
Remarks:
Since the :READ? query itself does not apply the trigger, the trigger is
required additionally. Once the :READ? query is executed, the device
does not execute the next command until the measured data is placed in
the output queue. For this reason, the trigger is not executed even if the
trigger source is set to BUS and an attempt is made to apply the trigger
by remote control immediately after the :READ? query, causing the
device to be hung up in the trigger waiting state. The trigger waiting
state can be reset by the device clear.
Содержание ZM2371
Страница 1: ...NF Corporation LCR METER ZM2371 ZM2372 Instruction Manual...
Страница 2: ......
Страница 3: ...LCR METER ZM2371 ZM2372 Instruction Manual DA00026021 007...
Страница 250: ......
Страница 252: ......
Страница 253: ......
Страница 254: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...
Страница 255: ...BLANK...