![NF ZM2371 Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/nf/zm2371/zm2371_instruction-manual_1684900061.webp)
3.5 Basic Operations
3-31
ZM2371/ZM2372
3.5.6 Setting of Basic Measurement Conditions
3.5.6.1 Measurement Frequency
The measurement frequency is displayed at the upper right of measurement screen.
・ ・ ・ ・ ・ ・ ・
1 . 0 0 0 0 k
Measurement frequency Hz
・ ・ ・ ・ ・ ・ ・
1 . 0 0 V
_
Measurement signal level
_
FREQ
_
Press the
_
FREQ
_
key to display the measurement frequency setting menu.
F r e q u e n c y : 1 . 0 0 0 0 k H z
Current setting
0 ) 1 2 0 1 ) 1 k 2 ) 1 0 k 3 ) 1 0 0 k 4 ) E n t r y
Options
120
120Hz
1k
1kHz
10k
10kHz
100k
100kHz
Entry
Displays the measurement frequency input menu to input a numeric
value.
If a setting is made or the EXIT operation is performed, the measurement screen comes
back.
Entry: The measurement frequency input menu is as follows.
F r e q u e n c y : 1 . 0 0 0 0 k H z
Current setting
0 . 0 0 1 H z t o 1 0 0 . 0 0 k H z
Range of settable values
The setting resolution is 5 digits (1mHz when < 10Hz).
If a setting is made or the EXIT operation is performed, one-previous menu comes back.
_
FREQ
_
Содержание ZM2371