![NF ZM2371 Скачать руководство пользователя страница 72](http://html1.mh-extra.com/html/nf/zm2371/zm2371_instruction-manual_1684900072.webp)
3.5 Basic Operations
3-42
ZM2371/ZM2372
3.5.6.5 Measurement Speed
_
SPEED
_
Press the
_
SPEED
_
key to display the measurement speed setting menu.
_
SPEED
_
M e a s u r i n g Sp e e d : FA S T
Current setting
0)RAPID 1 ) FA S T 2 ) M E D 3 ) S L O W 4 ) Ve r y S L O W
Range of settable values
If a setting is made or the EXIT operation is performed, one-previous menu comes back.
Current measurement speed is displayed on the second line of measurement screen.
A 1 0 0
FAST
2 5 6 I n t O p S h L d 1 m
・ ・ ・ ・ ・ ・
SPEED/AVERAGE
Measurement speed: RAPid / FAST / MEDium / SLOW / VerySLOw
(Lower-case letters are not displayed)
Table 3
–4 Example of measurement time (typical values)
Measurement
frequency
Measurement speed
RAP
FAST
MED
SLOW
VSLO
(
DC
)
148ms
148ms
148ms
215ms
613ms
120Hz
10ms
(1 cycle)
10ms
(1 cycle)
26ms
(3 cycles)
126ms
(15 cycles)
501ms
(60 cycles)
1kHz
2ms
(1 cycle)
5ms
(4 cycles)
25ms
(24 cycles)
121ms
(120 cycles)
501ms
(500 cycles)
10kHz
3ms
5ms
25ms
122ms
502ms
100kHz
3ms
5ms
25ms
122ms
502ms
Conditions: Measurement range hold, Trigger delay time = 0, Averaging count = 1
The measurement time of direct current (DC) is the time added when DC
resistance is measured.
The measurement time at other frequencies is the time from input of TRIG signal
to output of EOM signal through the handler interface when DC resistance is not
measured.
The values in ( ) at 120Hz and 1kHz express the signal acquisition time with the
cycles of the signal.
The signal acquisition time is obtained by subtracting the calculation time of about
1ms from the time given in table.
However, the signal acquisition time of direct-current resistance Rdc is obtained by
subtracting about 83ms from the time given in table.
For details of signal acquisition time
”Figure 3-6 Signal acquisition time”
After the signal acquisition, the DUT can be replaced.
Содержание ZM2371
Страница 1: ...NF Corporation LCR METER ZM2371 ZM2372 Instruction Manual...
Страница 2: ......
Страница 3: ...LCR METER ZM2371 ZM2372 Instruction Manual DA00026021 007...
Страница 250: ......
Страница 252: ......
Страница 253: ......
Страница 254: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...
Страница 255: ...BLANK...