4.2 Using Sequence Function
DP Series
75
Table 4-1 Wattage Limiter Value per Phase (Reference value, Unit: W)
Num
ber o
f power uni
ts
pe
r phas
e
Rated output
per phase
Single-phase model
Power Input
Single-phase three-wire model
Power Input
Three-phase model
Power Input
100 V system 200 V system 100 V system 200 V system 100 V system 200 V system
Note: Power input 100 V system is for power input voltage 170 V or less.
--------
Notes
----------------------------------------------------------------------------------------------------------------
If the wattage limiter operates, suppress the output wattage within the rated output power by
lowering the output voltage setting value.
Active power (unit: W) is restricted by the wattage limiter, however, reactive power (unit: var)
is not restricted. Therefore, the apparent power (unit: VA) may exceed the rated power
depending on the power factor of the load. Set the output voltage so that the apparent power
does not exceed the rated power, regardless of the operation of the wattage limiter.
The value of the wattage limiter cannot be changed. To limit the output power to a value that is
lower than the rating, decrease the number of power units that are energized through the power
unit energization setting, or use the peak current limiter or the RMS current limiter.
When the power input voltage is 170 V or less, the maximum output active power of a model
with the rated power of 6 kVA or higher is limited to 4.725 kW of maximum output active
power by the wattage limiter. However, for a single-phase three-wire model and a three-phase
model, the maximum output active power per phase is limited to 4.725 kW/2 and 4.725 kW/3.
---------------------------------------------------------------------------------------------------------------------------------
4.2 Using
Sequence
Function
4.2.1 Basics
By using Sequence function (optional), an output setting can be changed step-by-step according to
the program (sequence) created beforehand. This section describes the basics for using the Sequence
function.
AC/DC Mode and Signal Source
Sequence function keeps the selection of the AC/DC mode and a signal source, independently from
continuous output function/Simulation function. AC/DC mode can be selected from AC/ACDC/DC.
A signal source is fixed as INT.
Содержание DP015S
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Страница 4: ......
Страница 18: ......
Страница 19: ...DP Series 1 1 Outline 1 1 Overview 2 1 2 Series Line up and Polyphase System Configuration 2 1 3 Features 4 ...
Страница 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Страница 215: ...DP Series 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202 ...
Страница 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Страница 225: ...DP Series 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 214 ...
Страница 237: ...7 8 Replacement Air Filter DP Series 219 フロントグリル フロントグリル Front grill 1 Front grill 2 Figure 7 2 Front Grills ...
Страница 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Страница 239: ...DP Series 221 8 Troubleshooting 8 1 Error Messages and Error Handling 222 8 2 When a Failure is Suspected 229 ...
Страница 298: ......
Страница 300: ......
Страница 302: ......
Страница 303: ......
Страница 304: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN ...