4.1 Using Limiter, Setting Limiter
DP Series
71
2.
Set the value of RMS current limiter in [Io] item.
3.
Set the [Output OFF] item as [Enable] or [Disable]. When setting to [Enable], also set the time
from the start of limiter operation to the output off.
4.
Close the window.
4.1.3 Using Voltage Setting Range Limit
For the continuous output function, when the signal source is INT, SYNC, or ADD, the range of the
output voltage setting value to the internal signal source can be limited. Even if trying to set a
voltage that exceeds the limit, the setting value is restricted to the limiting value. This can prevent
mis-setting when you do not want to set values outside the range.
As for the voltage setting range limiting values, each value is kept independently for the following
settings for 100 V and 200 V ranges respectively: SIN/CLP waveform in AC mode, ARB waveform
in AC mode, ACDC mode, and DC mode.
In case of polyphase output, the setting of the voltage setting range limit is common to all phases,
and kept as the effective value of the phase voltage in this product. When the line voltage setting
(effective value setting) is selected in balanced mode, it is converted into the effective value setting
of the phase voltage according to the following formula and kept in this product.
Содержание DP015S
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Страница 4: ......
Страница 18: ......
Страница 19: ...DP Series 1 1 Outline 1 1 Overview 2 1 2 Series Line up and Polyphase System Configuration 2 1 3 Features 4 ...
Страница 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Страница 215: ...DP Series 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202 ...
Страница 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Страница 225: ...DP Series 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 214 ...
Страница 237: ...7 8 Replacement Air Filter DP Series 219 フロントグリル フロントグリル Front grill 1 Front grill 2 Figure 7 2 Front Grills ...
Страница 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Страница 239: ...DP Series 221 8 Troubleshooting 8 1 Error Messages and Error Handling 222 8 2 When a Failure is Suspected 229 ...
Страница 298: ......
Страница 300: ......
Страница 302: ......
Страница 303: ......
Страница 304: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN ...