4.3 Using Power Fluctuation Testing (Simulation) Function
DP Series
105
Relationship between each step and settable Step parameters
定常
1
移行
1
異常
移行
2
定常
2
T1
T2
T3
T4
T5
S1
S2
S4
S3
S5
トリガ出力
同期出力
V1, F1
V3, F3
V1, F1
電圧・周波数は
正常
1
と同じ
Voltage and
frequency are
same as
Normal 1
Trigger output
Sync output
Normal1
Normal2
Abnormal
Trans1
Trans2
Figure 4-11 Simulation Function Steps
Table 4-5 Simulation Function Steps and Settable Step Parameters
Step Initial
Normal 1
Trans 1
Abnormal
Trans 2
Normal 2
Step
time
— T1 T2 T3 T4 T5
AC
voltage
V0 V1 — V3 —
—(=V1)
Frequency
F0 F1 — F3 —
—(=F1)
Start phase
ON/OFF
P0
ON/OFF
P1
—
ON/OFF
P3
—
ON/OFF
P5
Stop Phase
ON/OFF
Q0
ON/OFF
Q1
—
ON/OFF
Q3
—
ON/OFF
Q5
Trigger
output —
ON/OFF ON/OFF ON/OFF ON/OFF ON/OFF
Sync code
output
S0 S1 S2 S3 S4 S5
Note: "—" means that this parameter cannot be set.
Start phase (Start Phs)
ON or OFF can be selected for the start phase. When ON is selected, the step starts at the specified
start phase. When OFF is selected, the step starts at the phase in which the previous step ended.
When selecting ON, a numerical entry box in which you can specify the start phase opens.
The start phase can be set at the Normal 1, Normal 2, and Abnormal steps. For other steps, the start
phase is fixed to OFF and cannot be specified.
Stop phase (Stop Phs)
ON or OFF can be selected for the stop phase. When ON is selected, the step ends at the specified
stop phase. If the phase is not the specified stop phase after the Step Time elapsed, the execution
continues the output of that step until reaching the specified Stop Phase, then performs the next step.
When OFF is selected, the step transitions to the next step when the Step Time has elapsed
regardless of the phase. When selecting ON, a numerical entry box in which you can specify the stop
Содержание DP015S
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Страница 4: ......
Страница 18: ......
Страница 19: ...DP Series 1 1 Outline 1 1 Overview 2 1 2 Series Line up and Polyphase System Configuration 2 1 3 Features 4 ...
Страница 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Страница 215: ...DP Series 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202 ...
Страница 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Страница 225: ...DP Series 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 214 ...
Страница 237: ...7 8 Replacement Air Filter DP Series 219 フロントグリル フロントグリル Front grill 1 Front grill 2 Figure 7 2 Front Grills ...
Страница 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Страница 239: ...DP Series 221 8 Troubleshooting 8 1 Error Messages and Error Handling 222 8 2 When a Failure is Suspected 229 ...
Страница 298: ......
Страница 300: ......
Страница 302: ......
Страница 303: ......
Страница 304: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN ...