10. Specifications
PROGRAMMABLE AC POWER SOURCE
266
10.26 External Control I/O
When you enabled the external control from the menu, you can control this product by using the
external signal (or non voltage contact). The state output is always on. The detection and state
switching is done at 10 ms-cycle. If the Remote state is achieved by the external interface, a control
input is ignored.
Item Description
External control input
Enabled or Disabled
Input level
TTL
Nondestructive maximum input
+10 V /
−
5 V
Input impedance
Pull-up to +5 V at 47 k
Ω
Output Off
Falling Off
Output On
Falling On
Sequence start/resume
Falling Start
Stop of sequence
Falling Stop
Sequence is in the Hold status
Falling Hold
Sequence branch 1
Falling Branch start
Sequence branch 2
Falling Branch start
Memory recall (+ compile)
Falling Recall
Memory specification 1
Memory specification 2
Specify 0 to 3
(Equivalent to memory 1 to 4,
respectively)
Control
input
Control
Clear the current peak-hold value
Falling Clear
Output level
TTL
Output impedance
220
Ω
Polarity
Polarity reverse (collective) is available
only for the state with *
Power On/Off
Low: Off, High: On
Output On/Off*
Low: On, High: Off
Protection operation*
Low: Active, High: None
Limiter operation*
Low: Active, High: None
AGC/Autocal setting state*
Low: On, High: Off
Software busy*
Low: Busy, High: Ready
←
Factory
default
(negative)
Output range
Low: 200 V, High: 100 V
Sequence operation
Step sync 1
Sequence operation
Step sync 2
High level or Low level
State
output
Status
Trigger positive
or negative
Terminal D-sub 25-pin multi-connector (rear panel)
Note: Input TTL level (
Low
: 0.8 V or lower, High: 2.6 V or higher)
Output TTL level (Low: 0.4 V or lower, High: 2.7 V or higher)
All means the chassis potential.
Note: The memory recall input of the control input recalls the setting memory for the Normal
(Continuous), the Sequence Memory for the Sequence, and the Simulation Memory for the
Simulation. For the Sequence and Simulation, the compile data is also included.
Note: Sequence start and stop of the control input are effective for the Simulation as well. Also,
step sync output 1 and 2 of the control output are effective for the Simulation as well.
Содержание DP015S
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Страница 4: ......
Страница 18: ......
Страница 19: ...DP Series 1 1 Outline 1 1 Overview 2 1 2 Series Line up and Polyphase System Configuration 2 1 3 Features 4 ...
Страница 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Страница 215: ...DP Series 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202 ...
Страница 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Страница 225: ...DP Series 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 214 ...
Страница 237: ...7 8 Replacement Air Filter DP Series 219 フロントグリル フロントグリル Front grill 1 Front grill 2 Figure 7 2 Front Grills ...
Страница 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Страница 239: ...DP Series 221 8 Troubleshooting 8 1 Error Messages and Error Handling 222 8 2 When a Failure is Suspected 229 ...
Страница 298: ......
Страница 300: ......
Страница 302: ......
Страница 303: ......
Страница 304: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN ...