10.15 Remote Sensing
DP Series
257
Voltage setting limit 2 (other than Voltage setting limit 1, phase voltage setting only)
All
models
Setting Range
(Peak value)
+10.0 V to +220.0 V / +20.0 V to +440.0 V
Positive
voltage
Factory
default
+220.0 V / +440.0 V
Setting Range
(Peak value)
−
220.0 V to
−
10.0 V /
−
440.0 V to
−
20.0 V
Negative
voltage
Factory
default
−
220.0 V /
−
440.0 V
Resolution 0.1
V
Note: The limitation is applied to the additional values of the AC voltage setting (recalculated to a
peak value) and the DC voltage setting.
Frequency setting limit (the lower limit
≤
the upper limit)
All
models
Setting Range
1.0 Hz (AC mode : 40 Hz) to 550.0 Hz
Upper
limit
Factory default
550.0 Hz
Setting Range
1.0 Hz (AC mode : 40 Hz) to 550.0 Hz
Lower
limit
Factory default
1.0 Hz (AC mode : 40 Hz)
Resolution 0.1
Hz
Note: In the AC mode, the setting range is 40.0 Hz to 550.0 Hz.
10.15 Remote Sensing
This switches the voltage used for measurement. When the remote sensing is on, the sensing input
terminal voltage is used. When it is off, the output terminal voltage is used.
By combining with AGC or Autocal, a voltage drop due to wiring to the load can be compensated.
When the remote sensing is on, the output voltage detection point corrected by the AGC or Autocal
function is switched to the sensing input terminal. When AGC or Autocal is off, only the detection
voltage used for measurement display is switched.
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the waveform
is sine wave or DC. It is turned off when the Sequence or Simulation is selected.
AGC/Autocal
Measurement
voltage, power, power factor
Off On
On
Use the sensing input terminal voltage
Not active
Active
Off
Use the output terminal voltage
Not active (factory
default)
Active
Содержание DP015S
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Страница 4: ......
Страница 18: ......
Страница 19: ...DP Series 1 1 Outline 1 1 Overview 2 1 2 Series Line up and Polyphase System Configuration 2 1 3 Features 4 ...
Страница 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Страница 215: ...DP Series 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202 ...
Страница 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Страница 225: ...DP Series 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 214 ...
Страница 237: ...7 8 Replacement Air Filter DP Series 219 フロントグリル フロントグリル Front grill 1 Front grill 2 Figure 7 2 Front Grills ...
Страница 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Страница 239: ...DP Series 221 8 Troubleshooting 8 1 Error Messages and Error Handling 222 8 2 When a Failure is Suspected 229 ...
Страница 298: ......
Страница 300: ......
Страница 302: ......
Страница 303: ......
Страница 304: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN ...