EDH0170FE1010 – 09/98
44
LDS1000 Controller for Laser Diode Sensors
10.5.05 RB
Read TRIGG Input Status
Use
h
MNL
j
RMT
j
ACT
Syntax RB
Parameter
None.
Description
Read the status of the TRIGG input. The return value will be
equal to 0 or 1 depending on whether the interface is in high
status or low status.
Response RBnn
nn = status of TRIGG input.
Errors
D Execution not allowed.
S Communication time-out.
Example
RB |
Request status of TRIGG input.
RB1
|
TRIGG input is in high status.
Related commands RO
Read PWOK output status.
TB
Read error.
TE
Read error code.
10.5.06 RD
Set Frozen Display Mode
Use
h
MNL
j
RMT
j
ACT
Syntax RD
Parameter
None.
Description
Freezes the display on “REMOTE MODE” instruction to remove
response times related to the display refresh.
This is the default display mode for Interface Mode.
Response
None.
Error
D Execution not allowed.
Example
RD |
Frozen display. The controller displays:
Related commands RE
Active display.
TB
Read error.
TE
Read error code.