47
EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.11 ST
Stop Work in Progress
Use
h
MNL
j
RMT
j
ACT
Syntax ST
Parameter
None.
Description
Stop work in progress (acquisition launched by TK, TM or TQ).
Response
None.
Error
D Execution not allowed.
Example
ST |
Stop the work in progress and carry on.
NOTE
The ST command stops the acquisitions in progress without deleting the
memory. The measurements performed up to receiving this command
are stored in the controller’s memory.
The XN parameter keeps its value.
The command xxTT makes it possible to read the values stored if xx is
lower than XN.
Related commands TB
Read error.
TE
Read error code.
TK
Acquisitions with trigger on TRIGG input.
TM
Immediate acquisitions.
TQ
Acquisitions with synchronization on TRIGG input.
TT
Proof read acquisitions.
XN
Read number of measurements in memory.