45
EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.07 RE
Set Active Display Mode
Use
h
MNL
j
RMT
j
ACT
Syntax RE
Parameter
None.
Description
Enable display in remote mode. The response times related to
display refresh may appear.
Response
None.
Error
D Execution not allowed.
Example
RE |
Active display.
Related commands RD
Frozen display.
TB
Read error.
TE
Read error code.
10.5.08 RO
Read PWOK Output Status
Use
h
MNL
j
RMT
j
ACT
Syntax RO
Parameter
None.
Description
Read the status of the PWOK output. The return value will be
equal to 0 or 1 depending on whether the interface is in high or
low status.
Response ROnn
nn = status of the PWOK output.
Errors
D Execution not allowed.
S Communication time-out.
Example
RO |
Request status of TRIGG input.
RO0
|
The PWOK output is in low status.
NOTE
The PWOK output is in status 1 if the sensor receives a sufficient quanti-
ty of light to carry out the measurement.
This logical output is available on the rear panel of the LDS1000 Controller.
Related commands TB
Read error.
TE
Read error code.