EDH0170FE1010 – 09/98
48
LDS1000 Controller for Laser Diode Sensors
10.5.12 TA
Query Sensor Parameters
Use
h
MNL
j
RMT
j
ACT
Syntax xxTA
Parameter xx
Axis number: 1 for Y and 2 for Z.
Possible errors
xx
≠
1 and xx
≠
2
☞
Error B.
Description
Read parameters of a current sensor axis:
TYPE : RT or TR.
COEF : 00100 to 29999.
INCR
: 0,1; 0,2; 0,5; 1; 2; 5; 10; 20; 50.
UNIT
: µrad; sec; mrad; nm; µm; mm; unit.
OFS1
: 00000 to 32767.
OFS2
: 00000 to 32767.
The parameters are separated by commas.
Response
Sensor axis parameters.
Errors
B Incorrect axis number.
D Execution not allowed.
S Communication time-out.
Example
1TA |
Read the parameters of axis 1 (Y) of the current sensor.
1TA,RT,04512,1,µrad,00102,00241
|
Typical parameters for an STR500-H autocollimator.
Related commands TB
Read error.
TE
Read error code.
10.5.13 TB
Error Description
Use
h
MNL
j
RMT
j
ACT
Syntax TBaa
Parameter aa
Error code in ASCII format (@ to U).
If there is no parameter
☞
read current error.
Description
Read message related to a an error code (parameter aa). Read
current error message if aa is missing.
Response TBaabb
aa = Error code queried.
bb = Error description.
Errors
D Execution not allowed.
S Communication time-out.
Example
TB@ |
Read error message @.
TB@ No error
|
No error.
Related command TE