49
EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.14 TE
Read Error Code
Use
h
MNL
j
RMT
j
ACT
Syntax TE
Parameter
None.
Description
Read last error code generated by the LDS1000 Controller:
ASCII character stored in error register.
The error register is reset when read with TB or TE.
Only the last error encountered can be read again.
Response TEaa
aa = Current Error Code.
Error messages:
@ : No error.
A : Unknown message code.
B : Incorect axis number.
C : Parameters out of limits.
D : Execution not allowed.
O : Variable is out of range.
S : Communication time-out.
Errors
D Execution not allowed.
S Communication time-out.
Example
TE |
Read error code.
TE@
|
No error.
Related command TB
Read error.