55
EDH0170FE1010 – 09/98
LDS1000 Controller for Laser Diode Sensors
10.5.22 VE
Read Internal Program Version
Use
h
MNL
j
RMT
j
ACT
Syntax VE
Parameter
None.
Description
Read internal program version of the controller.
Response VE LDS1000 Controller Version xx.yy
xx = Version.
yy = Progress index.
Errors
D Execution not allowed.
S Communication time-out.
Example
VE |
Read program version.
VE
LDS1000 Controller Version 3.10
|
Program version 3.10.
Related commands TB
Read error.
TE
Read error code.
10.5.23 XI
Read Integration Constant
Use
h
MNL
j
RMT
j
ACT
Syntax XI
Parameter
None.
Description
Read integration constant given in milliseconds.
Response XInn
nn is the integration constant given in milliseconds.
Errors
D Execution not allowed.
S Communication time-out.
Example
XI |
Read integration constant.
XI100
|
The integration constant is equal to 100 msec.
Related commands KI
Define the integration constant.
TB
Read error.
TE
Read error code.