LDS1000 Controller for Laser Diode Sensors
List of parameters:
7.4.2.3
Default Values of Communication Parameters
=>
(
F4
) =>
(
F2
)
The communication programs available for the LDS1000 Controller use the
communication parameter default values.
The LDS1000 Controller enables you to quickly return to the correct mode
of operation when the parameter modifications do not allow the Interface
Mode to be used normally.
For each communication mode, RS-232-C or IEEE-488, the
(
F3
) key
makes it possible to return to default values.
As a precaution, a confirmation messages asks if you really want to modify
the current parameters (and specifies the active interface):
7.4.3
LDS1000 Controller Parameter Input
=>
(
F4
) =>
(
F3
)
7.4.3.1
Reading the Internal Software Version
=>
(
F4
) =>
(
F3
)
This menu displays the version of the LDS1000 software installed in the
controller.
This internal software manages all the controller’s functions, in manual
mode and in Interface Mode. It also operates the controller when analog
outputs are installed. The physical modification of the software is done by
EPROM exchange on the parent card of the 68000 processor.
Should you encounter any operating problems with the controller, this
information will be useful to the maintenance service. It is possible to upda-
te the software as long as the material environment remains compatible
with the working of the software.
Access to Menu Rack provides the following display:
The following functions are active:
(
F4
)
Access to next parameter.
(
F5
)
Return to main menu.
Designation Available Values Default Value
MSG TERM
End of sequence CR; LF; CR LF; LF CR CR LF
terminator
PRIM ADRESS
IEEE-488 2 digits: 15
primary adress 00 to 30
SECOND ADRESS
IEEE-488 2 digits: 00
secondary adress 00 to 30
TIME OUT
Communication 0.01 to 10 sec 0.5
time-out