CHAPTER 4 PG-FP4 OPERATION USING GUI SOFTWARE
User’s Manual U15260EJ3V1UM
66
Figure 4-27. Setup Window – Supply Voltage
[Vdd [V]]
Displays the default Vdd level set by the parameter file in volts (V).
This Vdd level can be changed by input.
[Vdd2 [V]]
Displays the default Vdd2 level set by the parameter file in volts (V).
If the specifications of the target device require two types of Vdd for writing the flash memory, Vdd2 specifies
a low-level signal voltage. This voltage level can also be changed by input.
[On Target check box]
Check On Target if the V
DD
/V
DD2
voltages are supplied from the target system.
Caution
Set the values of the V
DD
/V
DD2
voltages supplied by the target system as Vdd [V] and Vdd2
[V] before checking On Target. The target system may be damaged if proper values are not
set.
[Vpp, Clk, Serial Port, Reset, H/S Input]
Display the default level of each pin set by the parameter file in volts (V).
The signal levels cannot be changed.