Operation
56
JuLI
TM
Stage - User Manual
Objective Lens
Provides the information which lens is currently used.
☞
Note: The field of view for the 4x, 10x, and 20x objectives is 3.55 mm², 0.58 mm²,
and 0.14 mm², respectively.
How to change the objective
1. Select an objective by clicking the blue radio button. This will cause the stage
to move to a neutral position.
2. Loosen and rotate the current objective counterclockwise to remove it from the
holder.
3. Place the new objective on the holder and rotate clockwise. Be sure to rotate
the objective until it is tight in the holder.
4. After replacing the objective, click
Set
to complete the routine.
☞
Note: Changing the lens without modifying the settings will result in
malfunctions of the device and the software.
☞
Note: The objectives are especially prone to damage by excessive ethanol.
Please always clean the objectives with ethanol-wetted objective cleaning
paper.
Image Output
The image output can be set to four different formats.
JPEG
,
BMP
, and
PNG
as 8-
bit and
TIFF
as 16-bit image.
Data Path
Allows users to assign a saving path and monitor the remaining hard drive volume
.