Ultrasonic milkanalyser
Operation manual
74/92
For pH electrodes, originally supplied with milk analyser read the
following information:
Temperature compensations
The output of pH electrodes varies with temperature in manner, predicted by
theory. When needed, Sensorex can supply electrode holders with build-in
automatic temperature compensators. The need of automatic compensation
depends on the temperature variation, the pH value being measured. At pH
of about 7 there is no error due to temperature and, of course, at a constant
temperature there is no error. As shown in the following table, the pH error
due to temperature is a function of both the temperature and the pH value
being measured. At a pH of about 7 there is no error due to temperature and,
of course, at a constant temperature there is no error. The more the
temperature changes from the ambient calibration temperature and the more
the pH departs from 7 the greater is the pH error.
pH temperature error table
C
pH
2
3
4
5
6
7
8
9
10
11
12
5
.30
.24
.18
.12
.06
0
.06
.12
.18
.24
.30
15
.15
.12
.09
.06
.03
0
.03
.06
.09
.12
.15
25
0
0
0
0
0
0
0
0
0
0
35
.15
.12
.09
.06
.03
0
.03
.06
.09
.12
.15
45
.30
.24
.18
.12
.06
0
.06
.12
.18
.24
.30
55
.45
.36
.27
.18
.09
0
.09
.18
.27
.36
.45
65
.60
.48
.36
.24
.12
0
.12
.24
.36
.48
.60
75
.75
.60
.45
.30
.15
0
.15
.30
.45
.60
.75
85
.90
.72
.54
.36
.18
0
.18
.36
.54
.72
.90
0 pH Error Range
Less than .1 pH Error Range
5. PH helpful hints
For greatest accuracy in pH measurement, follow these guidelines:
- Use the same technique to measure samples, which was used for
calibration.
- Be consistent with stirring rates, times and conditions.
- Calibrate with buffers, which are close in temperature to that of the
sample.
- Calibrate the pH electrode regularly, e.g. once a day for accuracy to
within 0.05 pH.