Ultrasonic milkanalyser
Operation manual
80/92
8. pH Calibration.
Fig. 13 Calibration buffers
Take out the sensor. You may not place the nut, because the analyzer do not
sucks during calibration.
Serves for calibrating the pH meter. For that purpose use 2 standard buffers,
shown on the display as Low buffer (for example 5.00 pH) and High buffer
(for example 7.00 pH). The procedure is following:
1.
Start menu
Calibration.
2.
Put the pH probe in Low buffer.
3.
By pressing the buttons of the analyser enter the exact buffer value. The
display shows:
pH Calibr
Put Izopot buff
Buf=xx.xxx