Ultrasonic milkanalyser
Operation manual
47/92
If the option
Large Disable
, is chosen, then only the main results are send to
the computer
– Fat, SNF, Density, Lac, Proteins, Added Water, sample
temperature, device serial number and calibration number.
If the option
Enable
, is chosen, except the above mentioned parameters also
data for Salts, Freezing Point, pH, Conductivity will be send to the computer. In
this case is necessary the software in the computer to be conformable to the
format of the sent data. After starting the menu, the display shows (for example):
Large Results
No
No OK Yes
7.2.3.4.2. Select High Fat
If the analyzer has embedded function for measurement of high fat products, by
this menu the calibration, with which this measurement to be made is chosen.
What is seen when this option is turned on is an obvious slowdown of the
sample’s suction speed.
7.2.3.4.3. HFSpeed for Cal
If the analyzer has embedded function for measurement of high fat products,
and a new calibration for high fat measurement is needed, before starting the
new calibration the operator has to start this menu. What is seen when this
option is turned on is an obvious slowdown of the sample’s suction speed during
calibration.
Please, pay attention to the fact that switching off the power supply cancels this
command action.
7.2.3.4.4. Reslt Precision
Serves for setting the precision of the measured results shown on the display. It
is set separately for every parameter, the precision could be 0.01 (standard) or
0.1.
7.2.3.4.5. PCB Main Identi
Gives information about the type and the version of the analyser (LS Identity).
7.2.3.4.6. Set Calibr Name.
Sets the names of the separate calibrations. The name could be chosen from
the group of predefined calibrations names or to edit a new one. When editing
the new name there is a possibility all ASCII codes to be used, as letters (caps
and normal), numbers and punctuation marks and popular symbols. The
calibration name consists of 8 symbols.
Содержание MIA-LA
Страница 1: ...Operation manual 1 92 MIA LA MILK ANALYZER LCD display 4 lines x 16 characters Operation manual...
Страница 12: ...Ultrasonic milkanalyser Operation manual 12 92 Fig 3 Principle working scheme...
Страница 13: ...Ultrasonic milkanalyser Operation manual 13 92 Fig 4 Connecting peripheral devices...