Ultrasonic milkanalyser
Operation manual
26/92
Always before the final acidic cleaning rinse the device with alkaline cleaning
solution by manual starting if it was not already automatically cleaned. It is
necessary to be done in order to remove the milk residues which could react
with the acidic cleaning solution.
Do not forget to rinse with clean water!
It is started using the working modes menu. After the measurement is
completed, by pressing and holding pressed the button
Enter
, the display
shows the possible working modes of the analyser.
Cow
Sheep
UHT
Cleaning
Final Clean
1.
Choose
Final Clean
. After choosing the final cleaning, the following
message appears on the display:
Put filled with
Water glass
and press Enter
to Continue
2. Put a glass filled with water.
3. The hose from the Waste pipe has to be placed in the vessel for waste
liquids.