Ultrasonic milkanalyser
Operation manual
41/92
which reminds us to make the next measurement. Before each measurement
the milk is stirred by pouring it 2-3 times from vessel to vessel. Continue the
procedure till the 5th measurement.
Cal 1 Samp Low
Fat = f.ff
SNF=y.yy
Den=d.dd
-------------------------
Lac=l.ll
Sol=s.ss
Pro=p.pp
Exit
After 5th measurement completion automatically appears the menu, which
reminds us to place the
Low fat
milk sample:
Cal: Cow
Put Sampl: Low
5 times
Stir 2-3 times the milk sample before each measurement by pouring it from
one vessel to another. The needed quantity is poured in the sample-holder
and it is put in the recess of the analyser. Start the measurement by pressing
the button
Enter
. The sample is sucked. Appears the following menu:
Make 5 times measurement of the low FAT sample.
After 5th measurement completion automatically appears the menu:
Cal: Cow
Put sample: Water
5 times
Which reminds for 5-times water measurement.
After the 5
th
measurement appears the menu:
Recalibrated
Power Off-On