Ultrasonic milkanalyser
Operation manual
29/92
8. After the cleaning procedure is finished, the following message appears,
reminding the the analyzer needs rinsing.
9. Rinsing procedure is executed, described in p. 2. After it is finished, the
display shows:
10. Place glass filled with acidic cleaning solution.
11. The pipe, coming from the orifice with inscription
с Waste Pipe has to be
placed in the glass.
12. Pressing button Enter starts the procedure for cleaning with acidic
solution.
The following appears on the display:
Where, X is changing number showing number of cycles done by the pump.
13. After the cleaning procedure is finished, the following message appears,
reminding the analyzer needs rinsing.
14. Rinsing procedure is executed, described in p. 2. After it is finished, the
display shows:
15. Procedure Final Clean is finished. Now the analyser is ready for normal
measurement or to be switched off from the power supply.
For maximum cleaning effect it is recommended the cleaning solution to be
preliminary heated up to 40-50 degrees centigrade.
Glass Acid Detrg
Waste pipe
Glass
press Enter
End of Cleaning
Ready to Start
Cleaning: X