Ultrasonic milkanalyser
Operation manual
45/92
Serve for assigning different working parameters (modes).
7.2.3.1. Net number.
Serves for assigning the device network number when connecting it in the
production network. The possible numbers are from 0 to 15 including.
After starting this function the display shows the following:
Net number
0
- OK +
By using the button “up”▲ the operator has the possibility to increase the
number, showing the channel’s number, and by button “down”▼, to decrease
it. Pressing the button
Enter
saves the chosen channel and exits the function.
When connected in the production network each device has to have a
unique number.
7.2.3.2.Recalibrate.
Serves for changing definite calibration. Methods are described in point 6.4.
7.2.3.3. Save/Rest Cal.
Through this menu you may save the new calibration in the device or to
restore the old one (factory) calibra
tion. This is necessary in case that you’ve
calibrated the device for cow milk, but after that the device is not measuring
correctly and you decide to restore the factory calibration settings. Position
the cursor across “
Restore calibration
” and press “
Enter
”
Possibilities:
Save calibration
– saves the chosen calibration in an internal buffer.
Restore calibration
– restores the chosen calibration from the internal
buffer.
The procedure
Save/Restore
is done for each calibration separately.